Additel 227
Documenting Multifunction Process Calibrator
Price on request
Additel’s new Multi-functional Documenting Process Calibrator series takes portability, functionality, and accuracy to a whole new level and packages it with an intuitive and easy to use color touchscreen display.
- Product description
- Extra information
- Video
- Downloads
-
Additel’s new Multi-functional Documenting Process Calibrator series takes portability, functionality, and accuracy to a whole new level and packages it with an intuitive and easy to use color touchscreen display. This series includes an advanced documenting pressure calibrator (ADT227) and an advanced documentation process calibrator with a built-in HART communicator (ADT227-HART). There is also a Profibus and Foundation Fieldbus module available. You need article code 9063 for this. Additionally, each calibrator has an ATEX certified intrinsically safe option (ADT227Ex) allowing you to perform calibration in the harshest of environments. We’re confident these new tools will not only meet your calibration requirements but will make metrology simple for you!
Also it is possible to connect external pressure modules from Fluke to the Additel 227 (non-Atex). You will need a special cable for this, article code 9062.
Recently it is also possible to place the ADT227 in a test bench, model ADT227P. With this we can deliver a test bench that is completely equipped with built-in Additel equipment. So you have all the functionality of a multifunctional calibrator in a test bench and in addition a mA input and 24V power supply for calibrating pressure transmitters in combination with an Additel pressure controller. -
Electrical Specifications - Source AccuracySpecificationsADT227
ADT227Ex RangeResolutionAccuracyRangeResolutionAccuracyVoltage DC0 to 15 V0.1 mV0.005%RDG+0.005%FS0 to 10.5 V0.1 mV0.01%RDG+0.005%FSCurrent DC0 to 25 mA0.1 uA0.01%RDG+0.005%FS0 to 25 mA0.1 uA0.01%RDG+0.005%FSResistance0 to 400 Ω1 mΩ0.005%RDG+0.005%FS0 to 400 Ω1 mΩ0.01%RDG+0.005%FS0 to 4000 Ω10 mΩ0.01%RDG+0.005%FS0 to 4000 Ω10 mΩ0.01%RDG+0.005%FSFrequency0.01 to 50000.0 HzAuto range, 6-digit0.002%RDG+2 on last digit0.01 to 50000.0 HzAuto range, 6-digit0.002%RDG+2 on last digitVoltage mV (TC)-10 to 75 mV0.1 uV0.0008%RDG + 0.004%FS-10 to 75 mV0.1 uV0.01%RDG + 0.004%FSPulse0 to 99999991N/A0 to 99999991N/AOptional rising edge and falling edge, minimum threshold voltage: 2.5V Loop power (max 25mA) 24VN/A±1 V20 VN/A± 10%
Electrical Specifications - Measurement Accuracy Specifications ADT227 ADT227Ex Range Resolution Accuracy Range Resolution Accuracy Voltage DC -300 to 300 mV 1 μV 0.005% RDG + 0.005% FS -300 to 300 mV 1μV 0.01% RDG + 0.005% FS -30 to 30 V 0.1 mV 0.005% RDG + 0.005% FS -30 to 30 V 0.1 mV 0.01% RDG + 0.005% FS Temperature Coefficient: ±5 ppm FS/°C (-10°C to 10°C and 30°C to 50°C) Impedance: -300 mV to 300 mV = > 100 MΩ
-30 V to 30 V = >1 MΩDC High Voltage -3 to 3 V 0.1 mV 0.05% RDG + 0.01% FS N/A -30 to 30 V 1 mV 0.05% RDG + 0.01% FS -300 to 300 V 10 mV 0.05% RDG + 0.01% FS Temperature coefficient: ±0.0025% FS/°C(-10°C to 10°C and 30°C to 50°C)Maximum input voltage = 300 V, IEC61010 300V CATII Commong mode rejection: >100 dB (at 50 or 60 Hz) Impedance: > 4 MΩ, DC coupling AC High Voltage 3V (40 to 500 Hz) 0.1 mV 0.5% RDG + 0.05% FS N/A 30V (40 to 500 Hz) 1 mV 0.5% RDG + 0.05% FS 300V (40 to 500 Hz) 10 mV 0.5% RDG + 0.05% FS Temperature coëfficiënt: ±(0.025% RD + 0.0025% FS)/°C
(-10°C to 10°C and 30°C to 50°C)Maximum input voltage = 300 V, IEC61010 300V CATII 9% to 100% of range is suitable for the above accuracy indicators Impedance: >4 MΩ, <100pF, AC coupling Current DC -30 to 30 mA 0.1 μA 0.01% RDG + 0.005% FS -30 to 30 mA 0.1 μA 0.01% RDG + 0.005% FS Temperature Coefficient: ±5ppm FS/°C (-10°C to 10°C and 30°C to 50°C), Impedance: < 40 Ω Resistance
(4-Wire)0 to 400 Ω 1 mΩ 0.005% RDG + 0.005% FS 0 to 400 Ω 1 mΩ 0.01% RDG + 0.005% FS 0 to 4000 Ω 10 mΩ 0.01% RDG + 0.005% FS 0 to 4000 Ω 10 mΩ 0.01% RDG + 0.005% FS Temperature coeficient: ±5 ppm FS/°C (-10°C to 10°C and 30°C to 50°C) 2-Wire + 50 mΩ, 3-wire+ 10 mΩ Excitation current: 0.25 mA Voltage mV (TC) -10 to 75 mV 0.1 μV 0.008% RDG + 0.004% FS -10 to 75 mV 0.1 μV 0.01% RDG + 0.004% FS Temperature Coefficient: ±5ppm FS/°C (-10°C to 10°C and 30°C to 50°C) Impedance: >100 MΩ Frequency 0.01 to 50000 Hz Auto range, 6-digit 0.002% RDG + 2 on last digit 0.01 to 50000 Hz Auto range, 6-digit 0.002% RDG + 2 on last digit Minimum threshold voltage: 2.5 V Supported units: Hz, kHz, MHz, CPM, CPH, s, ms, μs Pulse 0 to 9999999 1 N/A 0 to 9999999 1 N/A Optional rising edge and falling edge, minimum threshold voltage: 2.5V Switch Supports dry or wet switches. Voltage range of 3 to 30 V. Response speed < 10 ms General Specification Specifications ADT227 ADT227Ex Operating Temperature -10°C to 50°C -10°C to 50°C Specification guaranteed temperature range 10°C to 30°C 10°C to 30°C Storage Temperature -20°C to 70°C -20°C to 70°C Humidity <95%, non-condensing <95%, non-condensing Power supply 6600mAh, 23.8Wh lithium battery, charging time 4~6 hours, battery pack can be charged independently 4000mAh 14.4Wh Explosion-proof lithium battery packcharging time 6~8 hours, battery pack can be charged independently User interface Icon drive menus Icon driven menus with navigation buttons Ports protection voltage 50V max 30V max Display 5.0 inch 480 x 800 mm TFT LCD capacitive screen 4.4 inch 640 x 480 mm color display capacitive screen Maximum altitude 3000 meters European Compliance CE Mark Electrical Connection Ø4mm sockets and flat mini-jack thermocouple socket Size 6.97" x 4.13" x 2.04" (177 mm x 105 mm x 52 mm) Weight 1.6 lb (0.7 kg) 1.65 lb (0.75Kg) Battery Rechargeable Li-ion battery (included) Battery Life Typically 16 hours Typically 35 hours Battery Charge 110V/220V external power adapter included. Battery can be charged external to the unit. Typically charge time is 6-8 hours. External pressure module Dual channel serial plug, can connect two digital pressure modules Warm-up time Full specification performance is achieved after a 10 minute warm-up time. ROHS compliant Rohs II Directive 2011/65/EU, EN50581:2012 Display rate 3 readings per second Environmental parameter measurement Built-in barometer sensor (user-calibrated) IP protection level IP67, 1 meter drop test Communication Isolate USB-TYPEC (slave), Bluetooth BLE Calibration ISO 17025 accredited calibration with data Pressure Specification( ADT227 & ADT227Ex) The ADT161 and ADT161Ex series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 60,000 psi (-1 bar to 4200 bar). Accuracy from 0.02% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detailed specifications, please refer to the pressure modules datasheet. Temperature Specification Thermocouple Measurement and Source AccuracyADT227ADT227Ex TypeStandardTemperature Range (°C)Accuracy (°C)Standard
Temperature Range (°C)Accuracy (°C) Measure / Source Measure / Source SIEC 584-50 to 1768-50-00.76IEC 584-50 to 1768-50-1000.770-1000.56100-10000.42100-17680.441000-17680.47RIEC 584-50 to 1768-50-00.82IEC 584-50 to 1768-50-00.820-2000.570-2000.57200-17680.38200-17680.42BIEC 5840 to 1820200-3001.51IEC 5840 to 1820200-3001.51300-5001.00300-5001.00500-8000.62500-8000.62800-18200.43800-18200.43KIEC 584-270 to 1372-250 to -2000.72IEC 584-270 to 1372-250 to -2000.75-200 to -1000.23-200 to -1000.24-100 to 6000.12-100 to 6000.13600 to 13720.22600 to 13720.25NIEC 584-270 to 1300-250 to -2001.14IEC 584-270 to 1300-250 to -2001.17-200 to -1000.33-200 to -1000.34-100 to 13000.19-100 to 13000.22EIEC 584-270 to 1000-250--2000.39IEC 584-270 to 1000-250 to -2000.41-200--1000.15-200 to -1000.15-100-7000.09-100-7000.10700-10000.12700-10000.14JIEC 584-210 to 1200-210--1000.19IEC 584-210 to 1200-210 to -1000.20-100-7000.10-100-7000.11700-12000.15700-12000.17TIEC 584-270 to 400-250--1000.55IEC 584-270 to 400-250 to -1000.57-100-00.12-100-00.230-4000.080-4000.08CASTM E9880 to 23150 to 10000.24ASTM E9880 to 23150 to 10000.261000 to 18000.401000 to 18000.451800 to 23150.651800 to 23150.73DASTM E9880 to 23150-1000.31ASTM E9880 to 23150-1000.31100-12000.25100-12000.271200-20000.421200-20000.472000-23150.652000-23150.74GASTM E17510 to 231550-1000.90ASTM E17510 to 231550-1000.90100-2000.57100-2000.57200-4000.35200-4000.36400-15000.25400-15000.271500-23150.491500-23150.55LDIN43710-200 to 900-200 to -1000.11DIN43710-200 to 900-200 to -1000.12-100 to 4000.08-100 to 4000.09400 to 9000.10400 to 9000.12UDIN43710-200 to 600-200 to 00.21DIN43710-200 to 600-200 to 00.210 to 6000.080 to 6000.09Note: Internal CJC is ±0.15°C (-10°C to 50°C ambient temperature) Accuracy with external cold junction only, for internal cold junction add 0.15 ℃ (k=2) RTD Measurement and Source Accuracy
Measure and Simulate
Temperature Range (°C)
Accuracy (°C) ADT227
ADT227Ex PT10(385)
-200 to 850
-200~200
0.57
0.59
200~600
0.67
0.72
600~850
0.75
0.82
PT25(385)
-200 to 850
-200~200
0.24
0.27
200~600
0.30
0.35
600~850
0.34
0.41
PT50(3916)
-200 to 850
-200~200
0.13
0.16
200~600
0.17
0.22
600~850
0.20
0.27
PT100(385)
PT100(391)
PT100(3916)
PT100(3926)
-200 to 850
-200~200
0.08
0.10
200~600
0.11
0.16
600~850
0.14
0.20
PT200(385)
-200 to 850
-200~200
0.32
0.08
200~300
0.34
0.34
300~600
0.41
0.41
600~850
0.48
0.48
PT400(385)
-200 to 850
-200~0
0.15
0.04
0~200
0.18
0.18
200~600
0.25
0.25
600~850
0.30
0.30
PT500(385)
-200 to 850
-200~200
0.16
0.16
200~600
0.22
0.22
600~850
0.27
0.27
PT1000(385)
-200 to 850
-200~200
0.10
0.10
200~600
0.16
0.16
600~850
0.20
0.20
Cu10(427)
-200~260
-200~260
0.54
0.56
Cu50(428)
--200~260
-200~260
0.11
0.13
Cu100(428)
-200~260
-200~260
0.07
0.08
Ni100(617)
Ni100(618)
-60~180
-60~0
0.05
0.06
0~180
0.05
0.05
Ni120(672)
--80~260
-80~260
0.04
0.05
Ni1000
-50~150
-50~150
0.07
0.07
-
Documents:
Documents Upload Date Size Additel 227 Multifunction Process Calibrator Datasheet 2024.1.25 1.2 MB Additel 227 Documenting Process Calibrator User Manual 2022.2.10 2.0 MB Additel 226/227 Multifunction Process Calibrator Selection Guide 2022.6.30 0.26 MB 2024.1.30 0.4 MB Software and Firmware Name Versions Upload Date Size Additel 226/227 Firmware 1.0.0.7 2023-04-28 57 MB AMS driver 1.0.7 2022-11-28 12 MB Additel 226/227 Firmware Update Tool 0.33.29.7 2023-03-31 26 MB Additel 226/227 DD Library for Pressure 11.3.2022 2022-11-03 183 MB Additel 226/227 DD Library for Temperature 11.3.2022 2022-11-03 90 MB Additel 226/227 DD Library for Flow and Valves 11.3.2022 2022-11-03 325 MB