Additel 762
Automated Hydraulic Pressure Calibrators
Price on request
The Additel 762 Automated Pressure Calibrator is unlike any other pressure calibrator on the market.This revolutionary product is a complete turnkey solution for automation of pressure calibration work up to 1,000 bar (15,000 PSI).
- Product description
- Extra information
- Video
- Downloads
-
The Additel 762 Automated Pressure Calibrator is unlike any other pressure calibrator on the market.This revolutionary product is a complete turnkey solution for automation of pressure calibration work up to 15,000 PSI (1,000 bar).
Designed for use in both the field and the laboratory, the portability and accuracy of this state-of-the-art product will quickly become the favorite go-to calibrator for lab personnel and field technicians alike.
With fully automated support for calibration of pressure transmitters, switches, dial and digital gauges and sensors, including HART/PROFIBUS devices in conjunction with a fully integrated task feature, data collection and Wi-Fi connectivity, we had our customer’s needs in mind when designing our most capable pressure calibrator to date.
• Automated pressure generation and control to 15,000 psi (1,000 Bar)
• Accuracy to 0.01%FS
• Dual-range calibration
• Control stability ≤ 0.005%FS
• Portable - designed for use in the field and in the lab
• Control by external pressure modules
• Supports two external pressure modules
• Wi-Fi, LAN, Bluetooth, USB and Ethernet communication
• Full HART field communicator
• HART and profibus communication
• Data logging and task management
• Patented electric pump technology
• ISO 17025-accredited calibration w/data included -
Pressure Specifications Pressure Range 1-700 Bar (15-10,000 psi) and 1-1,000 Bar (15-15,000 psi) Range Selection Manual 3K psi, Manual 10K psi or Auto-range Accuracy GP15K: 0-15,000 psi, 0.01%FS(2)
GP15K:0-6,000 psi, 0.01%FS(2)GP15K: 0-15,000 psi, 0.01%FS(2)
GP15K:0-6,000 psi, 0.01%FS(2)GP10K: 0-10,000 psi, 0.01%FS(2)
GP10K:0-3,000 psi, 0.01%FS(2)GP10K: 0-10,000 psi, 0.01%FS(2)
GP10K:0-3,000 psi, 0.01%FS(2)Resolution 7 digits 6 digits Maximum External Load Capacity Max: 80 ml @ 700 Bar, 50 ml recommended Reservoir Max: 350 ml, built-in filter Control Stability (1) ≤ 0.005% FS from 100 to 10,000 psi Stability Duration > 5 min Pressure Modules Built-in one module with dual range External Control Pressure Module See the following "External Control Pressure Module Specification and Compatibility" table External Measurement Pressure Module All ADT161 pressure modules [1] Control Stability is based on the range selection or external pressure control module
[2] Pressure range capabilities are dependent on model. See ordering info.Electrical Specifications Specification Range Resolution Accuracy Note mA Measure -25 to 25 mA 0.1 μA ± (0.008 %RD + 1.0 μA) Impedance < 10 Ω -50 to 50 mA 0.1 μA ± (0.008 %RD + 2.0 μA) V Measure ±300 mV 1 μV ± (0.008 %RD + 6 μV) Impedance > 1 GΩ V Measure (Auto-ranging) -5 to 5 V 20 μV ± (0.008 %RD + 100 μV) Impedance > 1 MΩ -12 to 12 V 100 uV ± (0.008 %RD + 320 μV) -30 to 30 V 100 uV ± *0.008 %RD + 600 μV) Loop Power Source 24 V N/A ±1 V 50 mA (Max Loading) mA Source
0 to 2.5 mA 0.05μA 0.008 %RD + 0.2 μA 20 mA @ 1 kΩ 2.5 to 25 mA 0.5μA 0.008 %RD + 1.0 μA Power Source 16 to 30 V 1 V ±1 V 70 mA (Max Loading) V Source 0 to 16 V 0.25 mV 0.008 %RD + 500 μV Pressure Switch Mechanical Switch, Live Mechanical Switch, NPN Switch, PNP Switch N/A N/A Response time < 10 ms if the switch is powered. Voltage range: (3-30) V Temperature Compensation 18°C to 28°C Temperature Coefficient Outside of 18°C to 28°C: <±0.0005%RD+0.00005%FS/°C Misuse Protection Up to 30 V on any two sockets Pressure Switch Test Supported HART / Profibus PA Supported External Control Pressure Module Specification and Compatibility Specification Pressure Range Accuracy Pressure Type Media (psi) (bar) ADT161-01-GPXX for 0.01% FS 1000 70 0.01% FS Gauge G,L 1500 100 0.01% FS Gauge G,L 2000 140 0.01% FS Gauge G,L 3000 200 0.01% FS Gauge G,L 5000 350 0.01% FS Gauge G,L 10000 700 0.01% FS Gauge G,L 15000 1000 0.01% FS Gauge G,L ADT161-02-GPXX for 0.02% FS 1000 70 0.02% FS Gauge G,L 1500 100 0.02% FS Gauge G,L 2000 140 0.02% FS Gauge G,L 3000 200 0.02% FS Gauge G,L 5000 350 0.02% FS Gauge G,L 10000 700 0.02% FS Gauge G,L 15000 1000 0.02% FS Gauge G,L [1] For 0.01% FS pressure modules, single or multiple measurement range calibration certificate per request. For other specifications, please contact Additel for more details. General Specifications Specification Description User Interface Color touch screen and keypad operation Display 7” TFT touch screen 800 x 480 color Enclosure IP Rating IP31 Power Dedicated lithium battery or power adapter Battery Rechargeable Li-Ion battery, typically 12 hours of operation, less than 5 hours recharge. Weight 28 lbs (12.8 kg) without media Media Sebacate oil Size 11.81 x 8.66 x 7.56 in (300 x 220 x 192 mm) Communications Wi-Fi, LAN, Bluetooth, USB, Ethernet HART Communication Read, configure and calibrate HART devices - DD files updated periodically Data Storage > 8 GB Data Logging Up to 1,000,000 readings (data and time stamped) Task documentation Up to 1,000 tasks can be stored with data Automation Functions Switch test, auto step, leak test User Interface Localization English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian Pump life > 1,000,000 cycles Environmental Specifications Operating temperature: 32°F to 122°F( 0°C to 50°C) Storage temperature: -20°C to 60°C ( -4°F to 120°F) Humidity: <90%, non-condensing Certification ISO 17025 accredited certificate of calibration with NIST-traceable data Compliance CE Software Compatibility ACal, Additel Land and Additel Link for access via mobile application Warranty 1 year Hose & Filter End of Life The estimated End of Life (EOL) expectancy for all accessory hoses and filters (pneumatic and hydraulic) is approximately 10 years and should be replaced at the first sign of wear or damage.
-
Documents:
Documents Upload Date Size Additel 762 Automated Pressure Calibrator Datasheet 2022.5.5 1.3 MB Additel 762W Series Manual 2024.9.2 3.2 MB