Additel 850-1200 Laboratory Thermocouple Calibration Furnaces
Price on request
Thermocouple calibration work can be challenging. But with the Additel 850 that is history now.
- Product description
- Extra information
- Video
- Downloads
-
Thermocouple calibration work can be challenging.
At Additel, we understand the difficulties of this type of work. With the new ADT850 Laboratory Thermocouple Calibration Furnace is packed with features and a performance you will not find anywhere else. The ADT850 horizontal furnace covers a range of 300 to 1200°C and is commonly used in a multitude of industries such as energy, calibration laboratories, aerospace and metallurgy to name a few. It is generally used by primary and secondary calibration laboratories to calibrate noble and base metal thermocouples with the lowest possible uncertainties. Additel’s ADT850 is the most accurate, stable and versatile furnace available!• Temperature control from 300°C to 1200°C• Stability of ±0.1°C• Radial uniformity of ±0.25°C @ 1200°C• Axial uniformity of ±0.25°C @ 1200°C• Multi-zone temperature control• Portable, rugged and quick to temperature• Self-calibration feature (PC option)• Multi-zone temperature control• Quick cool technology• Sliding probe holder provides mechanical stability and precise probe depth control• Pivoting color touchscreen display• Internal and external probe control• Alumina and metal inserts available• Patent pending EMF shielding technology• Advanced safety control• Wi-Fi Communications -
Specification ADT850 Temperature Range 300°C to 1200°C Stability ±0.1°C/min Radial Uniformity ±0.25°C @ 1200°C Axial Uniformity ±0.25°C @ 1200°C Heating Time (23°C~1200°C) 40 mins, (empty well) Cooling Time (1200°C~300°C) 90 mins, (empty well) Operating Conditions 0°C to 50°C, 0-90%RH (0°C~50°C), non-condensing, <2000m altitude Storage Temperature -20°C to 70°C Display Screen 7 in (178 mm) color touch screen Display Resolution 0.01°C Heater Power 4000W (220V AC) Fuse T12A, 250V SIZE (W x H x L) 342 x 424 x 680 mm (13.5 x 16.7 x 26.8 in) Weight 45 kg (99.2 lbs) without insert Communication Wi-Fi, Bluetooth, USB, LAN Warranty 1 year AM1210-20-CJ Type S Reference Standard Thermocouple Temperature Range 0°C to 1300°C Type Type S: Platinum/10% Rhodium vs. platinum w/ cold junction Long Term Drift ±0.6°C at 1084.62°C after 1 year typical usage Short Term stability ±0.2°C at 1084.62°C Diameter of thermocouple wire 0.5 mm Sheath Material Quartz or Alumina Sheath Dimensions OD: 6 mm (0.236"); Length: 500mm (20.0") External Lead Wire S type thermocouple wire, 500 mm Protective Carrying Case Included Documentation Report of test with data -
Documents:
Filename Upload date Size Additel 850 Laboratory Thermocouple Calibration Furnace Datasheet 2021.1.28 0.74 MB Additel 850 Laboratory Thermocouple Calibration Furnace User Manual 2021.8.24 1.1 MB