Additel 878-1210 Thermocouple Calibration Furnaces
Price on request
At Additel they understand the many challenges associated with thermocouple calibration work. That is precisely why Additel decided to introduce the ADT875-1210 Thermocouple Calibration Furnace!
- Product description
- Extra information
- Video
- Downloads
-
At Additel they understand the many challenges associated with thermocouple calibration work. That is precisely why Additel decided to introduce the ADT875-1210 Thermocouple Calibration Furnace!
With an unmatched stability, uniformity and an optional on-board process calibrator, calibrating thermocouples has never been easier. With two separate units to choose from, the ADT875-1210 and ADT878-1210 furnaces include a patented multi-zone temperature control provides a never before seen, highly stable and uniform heat source to ensure you get the best possible results from a modest investment. With metallic interchangeable inserts, users have the flexibility needed to service a wide variety of UUT’s and the durability they have come to expect from Additel. The ADT875-1210 and ADT878-1210 can be purchased with or without our on-board process calibration electronics to provide flexibility for customers who are needing the best 1200°C heat source on the market.
If thermocouple calibration and/or verification work is part of your workload, you don’t want to miss out on this opportunity to save valuable time and money with these best in class furnaces from Additel.- Temperature control from 100°C to 1210°C
- Two models to choose from – Reference (ADT878) and Standard (ADT875)
- Stability of ±0.1°C
- 4 on-board measurement channels (PC option)
- Portable, rugged and quick to temperature
- Multi-zone temperature control
- Process calibrator option provides a multi-channel readout for TCs, switches and transmitters, including task documentation, and HART communications
- Internal and external sensor control
- Metallic interchangeable inserts
- Wi-Fi and Bluetooth capable
- Color touch screen display
- Self-calibration feature (PC option)
- Patent pending technology
-
General Specifications
Specification
875-1210
878-1210
Temperature Range
100°C to 1210°C
Display Accuracy
±1.2°C @ 100°C
±1.2°C @ 300°C
±1.2°C @ 600°C
±1.6°C @ 900°C
±2.0°C @ 1210°C
±1.0°C @ 100°C
±1.0°C @ 300°C
±1.0°C @ 600°C
±1.2°C @ 900°C
±1.5°C @ 1210°C
Stability
±0.1°C
Axial Uniformity (20mm zone)
±0.6°C @ 100°C
±0.9°C @ 300°C
±1.2°C @ 600°C
±1.5°C @ 900°C
±1.5°C @ 1210°C
±0.4°C @ 100°C
±0.6°C @ 300°C
±0.8°C @ 600°C
±1°C @ 900°C
±1°C @ 1210°C
Radial Uniformity
±0.2°C @ 100°C
±0.3°C @ 300°C
±0.4°C @ 600°C
±0.8°C @ 900°C
±1°C @ 1210°C
±0.2°C @ 100°C
±0.3°C @ 300°C
±0.4°C @ 600°C
±0.6C @ 900°C
±0.8°C @ 1210°C
Loading Effect
±0.5°C
Environmental Conditions
8°C to 38°C guaranteed accuracy
0°C to 50°C, 0% to 90% RH non-condensing, 3000 M altitude for normal operation
Storage Conditions
-20°C to 60°C
Immersion Depth
XR style inserts = 138 mm (5.43") XS style inserts = 116 mm (4.57")
(see insert ordering info for more details)
Insert Size - OD
24.8 mm (0.98 inches)
Heating Time
50 min: 23°C to 1210°C
Cooling Time
50 mins:1210°C to 300°C 50 mins: 300°C to 50°C
55 mins:1210°C to 300°C 55 mins: 300°C to 50°C
Typical Time to Stability
15 min
Resolution
0.01°C
Units
°C, °F, and K
Display
6.5 in (165 mm) color touch screen
Size (H x W x D)
345 x 170 x 330 mm (13.6 x 6.7 x 13.0 in)
Weight
10.6 kg (23.4 lbs)
Specification
875-1210
878-1210
Power Requirements
115 V AC (±10%) or 230 V AC (±10%),
switchable, 45-65 Hz, 580 W
Mechanical Testing
Vibration: 2 g (10-500 Hz), 30 min for 2 sides Impact: 4 g three times
Drop test: 500 mm (19.6 in)
Communication
USB A, USB B, RJ45, WiFi, Bluetooth
Localization
English, Chinese, Japanese, Russian, German
Warranty
1 year
Input Specifications (Process Calibrator [PC] Option)
Specification
875-1210
878-1210
TC Measurement Channels
Patented TC terminals: Accepting S, R, K, B, N, E, J, T, L, and U
TC Measurement Accuracy Type K
Ch. 1-4 (excluding sensor)
±0.182°C @ 100°C
±0.266°C @ 300°C
±0.310°C @ 600°C
±0.397°C @ 900°C
±0.517°C @1210°C
±0.172°C @ 100°C
±0.236°C @ 300°C
±0.251°C @ 600°C
±0.304°C @ 900°C
±0.382°C @ 1210°C
TC Range
–75 mV to 75 mV (UUT Channels 1-4)
–18 mV to 18 mV (Reference Channel)
TC Resolution
0.0001 mV, Input Impedance < 10Ω
TC Voltage Accuracy
0.02% RD + 8μV (ch. 1-4)
0.01% RD + 2μV (ref ch.)
0.01% RD + 8μV (ch. 1-4)
0.005% RD + 2μV (Ref ch.)
Internal CJC Accuracy
±0.35°C (ch. 1-4)
±0.25°C (ref ch.)
±0.30°C (ch. 1-4)
±0.20°C (ref ch.)
Current Range
–30 mA to 30 mA
Current Accuracy
±(0.02% of rdg+ 2μA)
±(0.01% of rdg + 0.6mV)
0.001 V; Input impedance:>
1M
0.0001V
0.0001V
DC 24V Output
24 V ± 10%, MAX 60 mA
Hart Communication
Optional (ADT875PC and ADT878PC Models)
"Temperature Coefficient 0°C to 8°C and 38°C to 50°C
TC Readouts: ±5 ppm FS/°C Current: ±5 ppm FS/°C Voltage: ±5 ppm FS/°C
Switch Test
Mechanical or Electrical - Channels 1 & 2 only
Documentation
Up to 1,000 tasks which store up to 10 results each containing as found and as left data. Snap shot feature allows for screen captures. Records auto step and ramp functions
TC Measurement Specifications and Calculations
(Process Calibrator [PC] Option)
TC Type
TEMP (°C)
Error (°C)[1]
TC Type
TEMP (°C)
Error (°C)[1]
875
878
875
878
K (CH1- CH4)
100
±0.182
±0.172
S (CH1- CH4)
100
±1.102
±1.094
300
±0.266
±0.236
300
±0.924
±0.899
600
±0.310
±0.251
600
±0.888
±0.837
900
±0.397
±0.304
900
±0.868
±0.793
1210
±0.517
±0.382
1210
±0.865
±0.765
N (CH1- CH4)
100
±0.273
±0.264
R (CH1- CH4)
100
±1.080
±1.072
300
±0.270
±0.243
300
±0.869
±0.844
600
±0.309
±0.256
600
±0.804
±0.755
900
±0.368
±0.285
900
±0.771
±0.699
1210
±0.455
±0.335
1210
±0.766
±0.670
E (CH1- CH4)
100
±0.136
±0.126
B (CH1- CH4)
250
±3.182
±3.170
300
±0.153
±0.130
300
±2.645
±2.631
600
±0.210
±0.154
600
±1.409
±1.379
900
±0.291
±0.202
900
±1.049
±1.003
1000
±0.297
±0.196
1210
±0.905
±0.839
L (CH1- CH4)
100
±0.223
±0.214
T (CH1- CH4)
100
±0.194
±0.185
300
±0.271
±0.241
300
±0.191
±0.166
600
±0.308
±0.251
400
±0.217
±0.183
900
±0.522
±0.448
S (EXT. REF)
100
±0.277
±0.273
U (CH1- CH4)
100
±0.270
±0.261
300
±0.242
±0.229
300
±0.189
±0.164
600
±0.249
±0.224
600
±0.227
±0.176
900
±0.258
±0.220
J (CH1- CH4)
100
±0.186
±0.177
1210
±0.266
±0.216
300
±0.197
±0.168
R (EXT. REF)
100
±0.271
±0.266
600
±0.256
±0.200
300
±0.228
±0.216
900
±0.281
±0.197
600
±0.227
±0.202
1200
±0.414
±0.294
900
±0.230
±0.194
1210
±0.240
±0.192
[1] Excluding cold junction compensation errors.
-
Documents:
Filename Upload date Size Additel 878 Thermocouple Calibration Furnaces Datasheet 2023.8.17 1.9 MB Additel 878 Thermocouple Calibration Furnaces User Manual 2020.5.19 1.3 MB
Update Packages:Filename Upload date Size Additel 878 Firmware Update Package 2021.11.18 11 MB