DPI 880
Price on request
The DPI 880 is an ultra-compact and simple to use multifunction calibrator for testing, configuring and calibrating virtually all process parameters. It is 35% smaller than competitive models and it?s windows-like menu driven display means fewer keys and a more simple, easier-to-understand interface. The DPI 880?s data logging capability means that users no longer need to carry additional recording tools.
- Product description
- Extra information
- Video
- Downloads
-
The DPI 880 is an ultra-compact and simple to use multifunction calibrator for testing, configuring and calibrating virtually all process parameters. It is 35% smaller than competitive models and it’s windows-like menu driven display means fewer keys and a more simple, easier-to-understand interface. The DPI 880’s data logging capability means that users no longer need to carry additional recording tools.
The DPI 880 Multi-function Calibrator is virtually a calibration lab in the palm of your hand. The DPI 880 allows the user to obtain high precision data and address any issue in as much time as it takes to press a button, saving thus valuable operating time. For example, the DPI 880
can accurately test switches, trips and alarms and print the results in 5 minutes, while a traditional device would take up to 40 minutes and would require manual data entry. Thermocouple transmitter calibrations
may take up to an hour using a conventional (older, more dated) tool that requires the use of look-up tables. The DPI 880 is capable of providing the ready to print results in just 5 minutes.Features
- Sources and reads mA, mV, ohms, frequency and pulses
- Simulates and reads 8 RTDs and 12 thermocouples
- 32 IDOS® ‘Plug-N-Play’ Pressure modules from 10 inH2O to 10,000 psi ( 25 mbar to 700 bar)
- Simultaneous dual reading capability
- Automatic switch test and pressure leak test
- 24 V loop power supply
- Data logging option
- Compact, simple to use, easy to carry
- Convenient one hand operation
Applications
- Test and maintenance
- Transmitter calibration
- Loop set-up and diagnostics
- Switch, trip and alarm verification
-
General specifications
Electrical connection
4mm sockets and mini-jack thermocouple socket
Calibrated temperature
50°F to 86°F (10°C to 30°C) unless otherwise stated
Operating Temperature
14°F to 122°F (-10°C to 50°C) unless otherwise stated Temperature coefficient 14°F to 50°F, 86°F to 122°F 0.0017%FS/°F (-10°C to 10°C, 30°C to 50°C 0.003%FS/°C) For ohms 14°F to 50°F, 86°F to 122°F 0.0028%FS/°F (-10°C to 10°C, 30°C to 50°C 0.005%FS/°C)
Storage Temperature
-4°F to 158°F (-20°C to 70°C)
Humidity
0% to 90% non-condensing, Def Stan 66-31, 8.6 Cat III
Shock and Vibration
BS EN61010:2001, Def Stan 66-31, 8.4 Cat III
EMC
BS EN61326-1:1998 + A2:2001
Safety
Electrical BS EN61010:2001. Pressure Equipment Directive (PED), Class SEP. CE marked
Display
Graphic LCD with backlight.
Size (l x w x h) and Weight
7.1 in x 3.3 in x 2 in (180 mm x 85 mm x 50 mm), 15 oz (425 g)
Batteries
3 AA alkaline, >50 hours measure, >10 hours 24V source
Electrical specification
Measure accuracy Source accuracy
0 to 24.000 mA
0.02% rdg + 2 counts
0 to 55.000 mA
0.02% rdg + 3 counts
0 to 120.00 mV
0.02% rdg + 2 counts
0 to 12.000 V
0.02% rdg + 2 counts
0 to 30.000 V
0.03% rdg + 2 counts
0 TO 4000.0 Ω
0.1 to 1.3 Ω
Switch detection
Open and closed, 2 mA current
Loop power
24 V +/-10% (35 mA maximum)
HART mA loop resistor
250 Ω (menu selection)
Frequency specification
Measure accuracy Source accuracy
0 to 999.999 Hz
0.003% rdg + 2 counts 0.003% rdg + 0.0023 Hz
0 to 50.0000 kHz
0.003% rdg + 2 counts 0.003% rdg + 0.0336 Hz
0 to 999999 cpm
0.003% rdg + 2 counts
0 to 59999 cpm
0.003% rdg + 0.138 cpm
0 to 999999 cph
0.003% rdg + 2 counts
0 to 99999 cph
0.003% rdg + 0.5 cph
Output waveform
Square wave (zero crossing)
Voltage input
30 V maximum
Trigger level
0 to 12 V resolution 0.1V
Output amplitude
0 to 12 VDC +/- 1% (10 mA maximum)
0 to 12 VAC pk-pk +/- 5% (10 mA maximum)
Temperature specification RTD Measure and Simulate
Standard
*Accuracy
Range
Pt 50 (385)
IEC 751
0.9°F (0.5°C)
-328°F to 1562°F (-200°C to 850°C)
Pt 100 (385)
IEC 751
0.45°F (0.25°C)
-328°F to 1562°F (-200°C to 850°C)
Pt 200 (385)
IEC 751
1.08°F (0.6°C)
-328°F to 1562°F (-200°C to 850°C)
Pt 500 (385)
IEC 751
0.72°F (0.4°C)
-328°F to 1562°F (-200°C to 850°C)
Pt 1000 (385)
IEC 751
0.36°F (0.2°C)
-328°F to 752°F (-200°C to 400°C)
D 100 (392)
JIS 1604-1989
0.45°F (0.25°C)
-328°F to 1202°F (-200°C to 650°C)
Ni 100
DIN 43760
0.36°F (0.2°C)
-76°F to 482°F (-60°C to 250°C)
Ni 120
MINCO 7-120
0.36°F (0.2°C)
-112°F to 500°F (-80°C to 260°C)
Ohms
0 to 4000
0.1 to 1.3 Ω
* Mid range figure quoted
Excitation: 0.2 to 0.5 mA measure, 0.05 to 3 mA simulate Pulse excitation currents minimum duration 10 ms
Temperature specification Thermocouple
Type
Standard
*Accuracy
Range
K
IEC 584
1.1°F (0.6°C)
-454°F to 2502°F (-270°C to 1372°C)
J
IEC 584
0.9°F (0.5°C)
-346°F to 2192°F (-210°C to 1200°C)
T
IEC 584
0.6°F (0.3°C)
-454°F to 752°F (-270°C to 400°C)
B
IEC 584
1.8°F (1.0°C)
32°F to 3308°F (0°C to 1820°C)
R
IEC 584
1.8°F (1.0°C)
-58°F to 3214°F (-50°C to 1768°C)
S
IEC 584
2.5°F (1.4°C)
-58°F to 3214°F (-50°C to 1768°C)
E
IEC 584
0.7°F (0.4°C)
-454°F to 1832°F (-270°C to 1000°C)
N
IEC 584
1.1°F (0.6°C)
-454°F to 2372°F (-270°C to 1300°C)
L
DIN 43710
0.6°F (0.3°C)
-328°F to 1652°F (-200°C to 900°C)
U
DIN 43710
0.6°F (0.3°C)
-328°F to 1112°F (-200°C to 600°C)
C
1.8°F (1.0°C)
32°F to 4199°F (0°C to 2315°C)
D
1.8°F (1.0°C)
32°F to 4514°F (0°C to 2490°C)
mV
0.2%rdg + 0.01%FS
-10 to 75 mV
*Mid range figure quoted
Cold Junction Error 0.4°F (0.2°C) maximum for 86°F (30°C) change in ambient temperature
-
Documents:
Filename Upload date Size DPI 880 Datasheet 2023.8.17 5.3 MB DPI880 Manual 2019.3.11 4.4 MB