DPI 800 / DPI 802
Price on request
The DPI 800 series is a complete series, advanced, robust and easy-to-use portable instruments. The tools are excellent for testing and calibrating of many process parameters. Advanced features and technical innovation make this instrument suitable for many applications and deliver results you can trust.
- Product description
- Extra information
- Downloads
-
The DPI 800 series is a complete series, advanced, robust and easy-to-use portable instruments. The tools are excellent for testing and calibrating of many process parameters. Advanced features and technical innovation make this instrument suitable for many applications and deliver results you can trust.
The DPI 800 Pressure Indicator is the ideal pressure measuring instrument, available with up to 2 internet pressure sensors. In addition, on the instrument called IDOS sensors are connected (intelligent pressure sensors that are recognized by the DPI 800 automtisch). You can easily extend in the future DPI 800 series. Available with pressure ranges from 25mbar to 700 bar including vacuum. Do you also want the oppertunity -besides measuring pressure- to measure signals and feed transmitters you could choose for the model DPI 802, the implementation that has a 24V loop power.Features
• Measuring ranges of 25 mbar to 700 bar
• Available with one or two internal sensors
• Pressure testing and maintenance
• Transmitter calibration
• 24V walking power
• Switch test functionApplications
• Testing and maintenance
• Transmitter calibration
• Loop test and diagnosis
• Switch test, trip and alarm verification
The DPI 800 series consists of a number of advanced, robust and simple to use portable instruments. These tools are ideal for testing and calibrating many popular process parameters. In addition, you can simply provide the tools to produce the so-called IDOS pressure sensors that are easy on one of the 800 DPI models can be connected, plug and play. -
DPI 802 Only
Measure
Accuracy
0 to 55.000 mA
0.02% reading + 3 counts
Temperature coefficient
14°F to 50°F , 86°F to 122°F, 0.0011%FS/°F (-10°C to 10°C, 30°C to 50°C, 0.002% FS/°C)
Switch detection
Open and closed. 2 mA current
Loop power output
24 V ±10% (35 mA maximum)
HART mA loop resistor
250 ∧ (menu selection)
Electrical connectors
4 mm sockets
Pressure Range G/D G A Media *Accuracy %FS + - S P ±10 in H2O (25 mbar) √ 0.1 0.03 ±1, 3, 5, or 10 psi
(70, 200, 350, or 700 m bar)
√0.075
0.03
5 psi (350 mbar)
√0.1
N/A
-15 to 15 or 30 psi
(-1 to 1 or 2 bar)
√0.05
0.01
30 psi (2 bar)
√ 0.075
N/A
-15 to 50, 100, 150 or
300 psi (-1 to 3.5, 7 10
or 20 bar)
√0.05
0.01
100, 300 psi (7, 20 bar)
√ 0.075
N/A
500, 1000, 1500, 2000
or 3000 psi
(35, 70, 100, 135, 200 bar)
√0.05
0.01
5000 or 10,000 psi
(350 or 700 bar) Sealed gauge
√0.05
N/A
G = gauge, A = absolute, G/D = gauge/differential; calibrated referenced to atmosphere maximum line pressure 30 psi (2 bar). Stainless steel, compatibility Non-corrosive gas/fluid and Non-corrosive gas. (N/A = not available). Accuracy assumes regular zero correction.
*S_Standard Accuracy
- Total accuracy over 32°F to 122°F (0°C to 50°C), including one year stability and calibration uncertainty
*P_Premier Accuracy
- Precision over 65°F to 82°F (18°C to 28°C)
- For operation over 41°F to 113°F (5°C to 45°C): 0.014% FS for ranges above 10 psi (700 mbar) 0.075% FS for ranges below 15 psi (1 bar)
- Stability over a year: 0.01% reading ranges above 5 psi (350 mbar), 0.03% reading ranges below 10 psi (700 mbar)
- Calibration uncertainty: 50 ppm of reading
-
Documents:
Filename Upload date Size DPI 800 / 802 Datasheet 2023.8.17 0.9 MB DPI 800/ 802 Manual 2019.3.11 0.8 MB